TY - JOUR
T1 - Pre-warming the Streamlined Liner of the Pharynx Airway (SLIPA™) improves fitting to the laryngeal structure
T2 - A randomized, double-blind study
AU - Kang, Hyun
AU - Kim, Dong Rim
AU - Jung, Yong Hun
AU - Baek, Chong Wha
AU - Park, Yong Hee
AU - Oh, Jong
AU - Kim, Won Joong
AU - Choi, Geun Joo
N1 - Publisher Copyright:
© 2015 Kang et al.
PY - 2015/11/20
Y1 - 2015/11/20
N2 - Background: The Streamlined Liner of the Pharynx Airway (SLIPA), a type of supraglottic airway, has a non-inflatable cuff that softens at body temperature to fit the laryngeal structure. We investigated whether pre-warming of SLIPA to body temperature may improve insertion parameters. Methods: Ninety adult patients were assigned equally randomized to either Group W or Group R. Anesthesia was induced using propofol, fentanyl, and rocuronium. In Group W, the SLIPA was warmed to 37 ° C before insertion, whereas in Group R, it was inserted at room temperature. The insertion time, oropharyngeal leak pressure, postoperative throat pain, blood staining, regurgitation, number of attempts at insertion, and difficulty of insertion were compared between the two groups. Results: The insertion time was shorter in Group W than in Group R (3.60 [3.15-4.06] s vs. 6.00 [4.45-7.50] s; P < 0.001). Oropharyngeal leak pressure from the time of insertion until 3 min after insertion was significantly higher in Group W than in Group R (P < 0.05). Postoperative throat pain, measured using the visual analog scale, was lower in Group W than in Group R (0.00 [0.00-2.50] vs. 2.00 [0.00-4.50]; P = 0.006). The difficulty of insertion was lower in Group W than in Group R (P < 0.004). There were no significant differences in terms of blood staining, regurgitation, and number of attempts. Conclusions: Pre-warming the SLIPA to body temperature has significant benefits compared to maintaining the device at room temperature. Specifically, insertion was easier, both insertion and fitting to the laryngeal structure could be performed more quickly, and the incidence of sore throat was reduced.
AB - Background: The Streamlined Liner of the Pharynx Airway (SLIPA), a type of supraglottic airway, has a non-inflatable cuff that softens at body temperature to fit the laryngeal structure. We investigated whether pre-warming of SLIPA to body temperature may improve insertion parameters. Methods: Ninety adult patients were assigned equally randomized to either Group W or Group R. Anesthesia was induced using propofol, fentanyl, and rocuronium. In Group W, the SLIPA was warmed to 37 ° C before insertion, whereas in Group R, it was inserted at room temperature. The insertion time, oropharyngeal leak pressure, postoperative throat pain, blood staining, regurgitation, number of attempts at insertion, and difficulty of insertion were compared between the two groups. Results: The insertion time was shorter in Group W than in Group R (3.60 [3.15-4.06] s vs. 6.00 [4.45-7.50] s; P < 0.001). Oropharyngeal leak pressure from the time of insertion until 3 min after insertion was significantly higher in Group W than in Group R (P < 0.05). Postoperative throat pain, measured using the visual analog scale, was lower in Group W than in Group R (0.00 [0.00-2.50] vs. 2.00 [0.00-4.50]; P = 0.006). The difficulty of insertion was lower in Group W than in Group R (P < 0.004). There were no significant differences in terms of blood staining, regurgitation, and number of attempts. Conclusions: Pre-warming the SLIPA to body temperature has significant benefits compared to maintaining the device at room temperature. Specifically, insertion was easier, both insertion and fitting to the laryngeal structure could be performed more quickly, and the incidence of sore throat was reduced.
KW - Oropharyngeal leak pressure
KW - Pre-warming
KW - SLIPA
KW - Supraglottic airway devices
UR - http://www.scopus.com/inward/record.url?scp=84947713681&partnerID=8YFLogxK
U2 - 10.1186/s12871-015-0151-4
DO - 10.1186/s12871-015-0151-4
M3 - Article
C2 - 26589142
AN - SCOPUS:84947713681
SN - 1471-2253
VL - 15
JO - BMC Anesthesiology
JF - BMC Anesthesiology
IS - 1
M1 - 167
ER -