TY - JOUR
T1 - Performance of the Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry system for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories
AU - Moon, Hee Won
AU - Lee, Sun Hwa
AU - Chung, Hae Sun
AU - Lee, Miae
AU - Lee, Kyungwon
PY - 2013
Y1 - 2013
N2 - We evaluated the performance of the Vitek MS for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories. With a total of 424 well-characterized isolates, the results of the Vitek MS were compared to those of conventional methods and 16S rRNA gene sequencing. The Vitek MS correctly identified 97.9% of the isolates tested to species level. The Vitek MS correctly identified the species of 97.2% of the staphylococci (95.9% of coagulasenegative staphylococci), 97.8% of the streptococci, and 100% of the enterococci. For the identification of Gram-positive cocci isolates, the overall concordance rate between conventional identification and the Vitek MS was 94.5 %. The Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system can be a reliable and rapid method for the identification of most relevant Gram-positive cocci. In addition, expanding the database of the Vitek MS, especially for coagulase-negative staphylococci, is needed to enhance the performance of the Vitek MS.
AB - We evaluated the performance of the Vitek MS for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories. With a total of 424 well-characterized isolates, the results of the Vitek MS were compared to those of conventional methods and 16S rRNA gene sequencing. The Vitek MS correctly identified 97.9% of the isolates tested to species level. The Vitek MS correctly identified the species of 97.2% of the staphylococci (95.9% of coagulasenegative staphylococci), 97.8% of the streptococci, and 100% of the enterococci. For the identification of Gram-positive cocci isolates, the overall concordance rate between conventional identification and the Vitek MS was 94.5 %. The Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system can be a reliable and rapid method for the identification of most relevant Gram-positive cocci. In addition, expanding the database of the Vitek MS, especially for coagulase-negative staphylococci, is needed to enhance the performance of the Vitek MS.
UR - http://www.scopus.com/inward/record.url?scp=84881600423&partnerID=8YFLogxK
U2 - 10.1099/jmm.0.062950-0
DO - 10.1099/jmm.0.062950-0
M3 - Article
C2 - 23764744
AN - SCOPUS:84881600423
SN - 0022-2615
VL - 62
SP - 1301
EP - 1306
JO - Journal of Medical Microbiology
JF - Journal of Medical Microbiology
ER -