Magnetic resonance imaging compatibility of the polymer-based cochlear implant

Jin Ho Kim, Kyou Sik Min, Soon Kwan An, Joon Soo Jeong, Sang Beom Jun, Min Hyoung Cho, Young Don Son, Zang Hee Cho, Sung June Kim

Research output: Contribution to journalArticlepeer-review

22 Scopus citations

Abstract

Objectives: In this study, we compared the magnetic resonance (MR) image artifacts caused by a conventional metal-based cochlear implant and a newly developed liquid crystal polymer (LCP)-based device. Methods: The metal-based cochlear implant system (Nurobiosys Co.) was attached to side of the head of a subject and the LCP-based device was attached to opposite side. In both devices, alignment magnets were removed for safety. Magnetic resonance imaging (MRI) was performed on a widely used 3.0 T and an ultra-high 7.0 T MRI machine. 3.0 and 7.0T MR images were acquired using T1- and T2 *-weighted gradient echo sequences, respectively. Results: In the 3.0 T images, the metal-based device on the left side generated the significant amount of artifacts. The MR images in the proximity of the metal package were obscured by the artifacts in both axial and sagittal views. On the other hand, the MR images near the LCP-based device were relatively free from the artifacts and clearly showed the brain structures. 7.0 T MR images showed the more severe distortion in the both sides but the metal-based cochlear implant system caused a much larger obscure area than the LCP-based system. Conclusion: The novel LCP-based cochlear implant provides a good MRI compatibility beyond present-day cochlear implants. Thus, MR images can be obtained from the subjects even with the implanted LCP-based neural prosthetic systems providing useful diagnostic information. Furthermore, it will be also useful for functional MRI studies of the auditory perception mechanism after cochlear implantations as well as for positron emission tomography-MRI hybrid imaging.

Original languageEnglish
Pages (from-to)S19-S23
JournalClinical and Experimental Otorhinolaryngology
Volume5
Issue numberSUPPL. 1
DOIs
StatePublished - 2012

Keywords

  • Cochlear implant
  • MRI compatibility
  • Magnetic resonance imaging

Fingerprint

Dive into the research topics of 'Magnetic resonance imaging compatibility of the polymer-based cochlear implant'. Together they form a unique fingerprint.

Cite this