TY - JOUR
T1 - Is follow-up testing with the filmarray gastrointestinal multiplex PCR panel necessary?
AU - Park, Sholhui
AU - Hitchcock, Matthew M.
AU - Gomez, Carlos A.
AU - Banaei, Niaz
PY - 2017/4
Y1 - 2017/4
N2 - The FilmArray gastrointestinal (GI) panel (BioFire Diagnostics, Salt Lake City, UT) is a simple, sample-to-answer, on-demand, multiplex, nucleic acid amplification test for syndromic diagnosis of infectious gastroenteritis. The aim of this study was to measure the yield of follow-up testing with FilmArray GI panel within 4 weeks of an initial test. Consecutive adult and pediatric patients tested at an academic institution between August 2015 and June 2016 were included in this study. Of 145 follow-up tests in 106 unique patients with an initial negative result, 134 (92.4%) tests and 98 (92.5%) patients remained negative upon follow-up testing. Excluding targets that are not reported at this institution (Clostridium difficile, enteroaggregative Escherichia coli, enteropathogenic E. Coli, and enterotoxigenic E. Coli), 137 (94.5%) follow-up tests and 101 (95.3%) patients remained negative. Weekly conversion rates were not significantly different across the 4-week follow-up interval. No epidemiological or clinical factors were significantly associated with a negative to positive conversion. Of 80 follow-up tests in patients with an initial positive result, 43 (53.8%) remained positive for the same target, 34 (42.5%) were negative, and 3 were positive for a different target (3.8%). Follow-up testing with FilmArray GI panel within 4 weeks of a negative result rarely changed the initial result, and the follow-up test reverted to negative less than half the time after an initial positive result. In the absence of clinical or epidemiological evidence for a new infection, follow-up testing should be limited and FilmArray GI panel should not be used as a test of cure.
AB - The FilmArray gastrointestinal (GI) panel (BioFire Diagnostics, Salt Lake City, UT) is a simple, sample-to-answer, on-demand, multiplex, nucleic acid amplification test for syndromic diagnosis of infectious gastroenteritis. The aim of this study was to measure the yield of follow-up testing with FilmArray GI panel within 4 weeks of an initial test. Consecutive adult and pediatric patients tested at an academic institution between August 2015 and June 2016 were included in this study. Of 145 follow-up tests in 106 unique patients with an initial negative result, 134 (92.4%) tests and 98 (92.5%) patients remained negative upon follow-up testing. Excluding targets that are not reported at this institution (Clostridium difficile, enteroaggregative Escherichia coli, enteropathogenic E. Coli, and enterotoxigenic E. Coli), 137 (94.5%) follow-up tests and 101 (95.3%) patients remained negative. Weekly conversion rates were not significantly different across the 4-week follow-up interval. No epidemiological or clinical factors were significantly associated with a negative to positive conversion. Of 80 follow-up tests in patients with an initial positive result, 43 (53.8%) remained positive for the same target, 34 (42.5%) were negative, and 3 were positive for a different target (3.8%). Follow-up testing with FilmArray GI panel within 4 weeks of a negative result rarely changed the initial result, and the follow-up test reverted to negative less than half the time after an initial positive result. In the absence of clinical or epidemiological evidence for a new infection, follow-up testing should be limited and FilmArray GI panel should not be used as a test of cure.
KW - FilmArray gastrointestinal
KW - GI PCR
KW - Repeat
KW - Syndromic
UR - http://www.scopus.com/inward/record.url?scp=85016200427&partnerID=8YFLogxK
U2 - 10.1128/JCM.02354-16
DO - 10.1128/JCM.02354-16
M3 - Article
C2 - 28122874
AN - SCOPUS:85016200427
SN - 0095-1137
VL - 55
SP - 1154
EP - 1161
JO - Journal of Clinical Microbiology
JF - Journal of Clinical Microbiology
IS - 4
ER -