Abstract
Aim: To evaluate the accuracy of depth measurements on supine chest computed tomography (CT) for transthoracic needle biopsy (TNB). Materials and methods: We measured skin-lesion depths from the skin surface to nodules on both prebiopsy supine CT scans and CT scans obtained during cone beam CT-guided TNB in the supine (n = 29) or prone (n = 40) position in 69 patients, and analyzed the differences between the two measurements, based on patient position for the biopsy and lesion location. Results: Skin-lesion depths measured on prebiopsy supine CT scans were significantly larger than those measured on CT scans obtained during TNB in the prone position (p < 0.001; mean difference ± standard deviation (SD), 6.2 ± 5.7 mm; range, 0-18 mm), but the differences showed marginal significance in the supine position (p = 0.051; 3.5 ± 3.9 mm; 0-13 mm). Additionally, the differences were significantly larger for the upper (mean ± SD, 7.8 ± 5.7 mm) and middle (10.1 ± 6.5 mm) lung zones than for the lower lung zones (3.1 ± 3.3 mm) in the prone position (p = 0.011), and were larger for the upper lung zone (4.6 ± 5.0 mm) than for the middle (2.4 ± 2.0 mm) and lower (2.3 ± 2.3 mm) lung zones in the supine position (p = 0.004). Conclusions: Skin-lesion depths measured on prebiopsy supine chest CT scans were inaccurate for TNB in the prone position, particularly for nodules in the upper and middle lung zones.