Abstract
Background: Temperature monitoring during high-intensity focused ultrasound (HIFU) therapy on tissue is essential to regulate the degree of thermal coagulation and to achieve the desired treatment outcomes eventually. The aim of the current study was to design and investigate the feasibility of a proportional-integral-derivative (PID) temperature controller-integrated portable HIFU driver for thermal coagulation. Methods: A portable HIFU driver was designed and operated at a maximum output voltage of 50 V with pulse-width modulation signals at 2 MHz. The temperature of ex vivo bovine liver tissue was monitored using a K-type thermocouple during the 2-MHz HIFU exposure. Results: The tissue temperature was maintained at 60 °C using a PID controller-integrated HIFU driver that modulated the output voltage during the 300-s HIFU exposure. The ex vivo testing demonstrated that the tissue temperature at the focal point approached the chosen temperature, i.e., 60 °C, within 70 s. The temperature was maintained with a deviation of less than 4 °C until the HIFU driver voltage was turned off at 300 s. Conclusions: The designed PID controller-integrated HIFU driver can be used as a small portable tool to regulate the tissue temperature in real time and achieve thermal coagulation via HIFU sonication.
Original language | English |
---|---|
Article number | 77 |
Journal | BioMedical Engineering Online |
Volume | 18 |
Issue number | 1 |
DOIs | |
State | Published - 26 Jun 2019 |
Bibliographical note
Publisher Copyright:© 2019 The Author(s).
Keywords
- Coagulation
- High-intensity focused ultrasound (HIFU)
- Proportional-integral-derivative (PID) controller
- Temperature control